Screening reactions in one-second DESI-MS droplets

Wleklinski M, Loren BP, Ferreira CR, et al. · Chemical science · 2018

Open access · cc by · source: Europe PMC

Desorption electrospray ionization mass spectrometry reads amine alkylation and Suzuki hits from nanoliter spots in about a second.

Key findings

Product ions appear within about 1 s per droplet. Benzyl bromide alkylated 7 of 8 amines versus 3 of 8 for 2-(bromoethyl)benzene. Peak analysis reached 6004 reactions per hour at 6144-well density, using less than 0.5 μg of each reagent in 50 nL spots.

Methodology

Authors spotted nanoliter reaction mixtures on porous PTFE, analyzed them by DESI-MS/MS, and screened N-alkylations plus Pd-catalyzed Suzuki couplings across multi-thousand-spot plates.

Limitations

Qualitative reproducibility falls at the highest speeds, and translating every DESI hit to optimized continuous-flow yield is not completed in this paper.